Case Reports[Removal of a swallowed endotracheal tube in a neonate].
Immediately following delivery a newborn infant was severely depressed. Because no respiratory effort was made and copious mucus was present, a Cole endotracheal tube was easily inserted into the trachea. Because of a mistaken manoeuvre for fastening the tube, the plastic adapter connection became dislodged from the tube, and the tube slipped out of trachea. ⋯ The misplaced tube was removed from the oesophagus using a small, flexible biopsy forceps closed, advanced into the lumen of the swallowed tube under fluoroscopic vision. By opening the forceps, gripping the tube tightly from within, it was easily withdrawn. No adverse effects were observed.
-
Minerva anestesiologica · Apr 1994
Case Reports[Intracardial knotting of a Swan-Ganz catheter. Description of a clinical case].
A 79 year old female was admitted to CTO Intensive Care Unit in the immediate postoperative period of orthopedic surgery. A Swan-Ganz fiber optic catheter was inserted through the right internal jugular vein but it was not possible to obtain a satisfactory tracing off pulmonary artery occlusion pressure. ⋯ It was possible to extract the knotted catheter through original venotomy so avoiding surgery. We want to emphasize that it is of fundamental importance in the insertion of the Swan-Ganz catheter to respect the recommended distances and to avoid repeated attempts to advance and withdraw the catheter.