Mycobacterium chimaera, a non-tuberculous mycobacterium, was recently identified as causative agent of deep-seated infections in patients who had previously undergone open-chest cardiac surgery. Outbreak investigations suggested an aerosol-borne pathogen transmission originating from water contained in heater-cooler units (HCUs) used during cardiac surgery. Similar thermoregulatory devices are used for extracorporeal membrane oxygenation (ECMO) and M. chimaera might also be detectable in ECMO treatment settings. ⋯ Among 118 ECMO patients, 76 had bronchial specimens analysed for mycobacteria and M. chimaera was found in three individuals without signs of mycobacterial infection at the time of sampling. We conclude that M. chimaera can be detected in water samples from ECMO-associated thermoregulatory devices and might potentially pose patients at risk of infection. Further research is warranted to elucidate the clinical significance of M. chimaera in ECMO treatment settings.
-
Following the reported link between heater-cooler unit (HCU) colonisation with Mycobacterium chimaera and endocarditis, mycobacterial sampling of all HCUs in use in Western Australia was initiated from August 2015, revealing M. chimaera colonisation in 10 of 15 HCUs. After M. chimaera was isolated from a pleural biopsy from a cardiothoracic patient who may have been exposed to a colonised HCU, a whole genome sequencing investigation was performed involving 65 specimens from 15 HCUs across five hospitals to assess if this infection was related to the HCU. Genetic relatedness was found between the 10 HCU M. chimaera isolates from four hospitals. However the M. chimaera isolate from the cardiothoracic patient was not genetically related to the HCU M. chimaera isolates from that hospital, nor to the other HCU isolates, indicating that the HCUs were not the source of the infection in this patient.