Journal of forensic sciences
-
The performance of the Microgenics CEDIA DAU assays for screening amphetamines, barbiturates, benzodiazepines, cocaine, opiates, phencylidine (PCP), and tetrahydrocannabinol (THC) was evaluated on the Boehringer Mannheim/Hitachi 717 in urine. Limits of detection ranged from 0.6 ng/mL for PCP, to 34.1 ng/mL for benzodiazepines. The average within run and total precision for these assays ranged from 1.3 to 7.3% for controls at cutoff concentrations, and control values at -25% and +25% of cutoffs. ⋯ CEDIA produced strong interferences for most drug assays in the presence of glutaraldehyde, detergent, and high concentrations of bleach and Drano. Minimal or selective interferences were seen with golden seal tea lemon juice, Visine, and low concentrations of bleach and Drano. Essentially no interference was observed with bicarbonate, sodium chloride, and vinegar.