The Journal of pediatrics
-
The Journal of pediatrics · Sep 1985
Comparative StudyPulse oximetry in pediatric intensive care: comparison with measured saturations and transcutaneous oxygen tension.
We evaluated a new pulse oximeter designed to monitor beat-to-beat arterial oxygen saturation (SaO2) and compared the monitored SaO2 with arterial samples measured by co-oximetry. In 40 critically ill children (112 data sets) with a mean age of 3.9 years (range 1 day to 19 years), SaO2 ranged from 57% to 100%, and PaO2 from 27 to 128 mm Hg, heart rates from 85 to 210 beats per minute, hematocrit from 20% to 67%, and fetal hemoglobin levels from 1.3% to 60%; peripheral temperatures varied between 26.5 degrees and 36.5 degrees C. ⋯ Simultaneous measurements with a tcPO2 electrode showed a similarly good correlation with PaO22 (r = 0.91), but the differences between the two measurements were much wider (mean 7.1 +/- 10.3 mm Hg, range -14 to +49 mm Hg) than the differences between pulse oximeter SaO2 and measured SaO2 (1.5% +/- 3.5%, range -7.5% to -9%) and were not predictable. We conclude that pulse oximetry is a reliable and accurate noninvasive device for measuring saturation, which because of its rapid response time may be an important advance in monitoring changes in oxygenation and guiding oxygen therapy.