Autoregressive modeling with exogenous input of middle-latency auditory-evoked potentials to measure rapid changes in depth of anesthesia.
Obtaining an adequate depth of anesthesia is a continuous challenge to the anesthetist. With the introduction of muscle-relaxing agents the traditional signs of awareness are often obscured, or difficult to interpret. These signs include blood pressure, heart rate, pupil size, etc. ⋯ The method was clinically evaluated in 10 patients anesthetized with alfentanil and propofol. The time interval between propofol induction and the time when the Na-Pa amplitude was decreased to 25% of the initial amplitude was measured. These measurements showed that ARX-estimated compared to MTA-estimated AEP was significantly faster in tracing transition from consciousness to unconsciousness during propofol induction (p < 0.05).