The Journal of thoracic and cardiovascular surgery
-
J. Thorac. Cardiovasc. Surg. · Sep 1989
Cardiopulmonary bypass without systemic heparinization. Performance of heparin-coated oxygenators in comparison with classic membrane and bubble oxygenators.
Performance characteristics of heparin-coated hollow-fiber membrane oxygenators (COATED HFMO, n = 5) were evaluated in an open-chest dog model without systemic heparinization. Four other oxygenators were evaluated with standard systemic heparinization (300 IU/kg, activated clotting time more than 400 seconds): a standard hollow-fiber membrane oxygenator (HFMO, n = 5), an inversed hollow-fiber membrane oxygenator (IHFMO, n = 5), a plate membrane oxygenator (PLATE MO, n = 5) and a bubble oxygenator (BUBBLE O, n = 5). The 25 dogs (36 +/- 12 kg) were perfused after cavo-aortic cannulation for 6 hours with a mean flow of 100 ml/kg body weight. ⋯ Determination of fibrin split products during perfusion without systemic heparinization did not show a significant increase. At the end of perfusion all devices were disconnected and gently rinsed with saline: There were no macroscopic clots in the COATED HFMO group perfused without systemic heparin. However, uncoated equipment introduced for control in animals perfused without systemic heparin showed major clotting.(ABSTRACT TRUNCATED AT 400 WORDS)