Ann Oto Rhinol Laryn
-
Ann Oto Rhinol Laryn · Sep 2002
Comparative StudyComparison of various methods of electromyographic monitoring of the recurrent laryngeal nerve in thyroid surgery.
In the present study, 3 types of electrodes for recurrent laryngeal nerve (RLN) monitoring are compared: 1) intralaryngeal surface electrodes attached to a conventional endotracheal tube, 2) monopolar needle electrodes placed on the vocal cords by direct laryngoscopy, and 3) bipolar needle electrodes inserted intraoperatively through the cricothyroid ligament. Data were collected from stimulation of 21 RLNs in 16 patients undergoing thyroid surgery. ⋯ The mean (+/-SD) amplitudes of the EMG potentials were 1.61 +/- 1.6 mV, 2.37 +/- 1.8 mV, and 0.35 +/- 0.4 mV for the monopolar endolaryngeal, bipolar transligament, and intralaryngeal surface electrodes, respectively. The advantages and disadvantages of each type of electrode are discussed.