Monitoring of patient-ventilator interactions at the bedside involves evaluation of patient breathing pattern on ventilator settings. One goal of mechanical ventilation is to have ventilator-assisted breathing coincide with patient breathing. ⋯ The types of asynchronies discussed are trigger asynchrony (ie, breath initiation that may manifest as ineffective triggering, double-triggering, or auto-triggering); flow asynchrony (ie, breath-delivery asynchrony, which may manifest as assisted-breath delivery being faster or slower than what patient desires); and cycling asynchronies (ie, termination of assisted inspiration does not coincide with patient breath termination, which may manifest as delayed cycling or premature cycling). Various waveforms are displayed and graphically demonstrate asynchronies; basic principles of waveform interpretation are discussed.
-
Patient-ventilator interaction is a key element in optimizing mechanical ventilation. The change from inspiration to expiration is a crucial point in the mechanically ventilated breath, and is termed "cycling." Patient-ventilator asynchrony may occur if the flow at which the ventilator cycles to exhalation does not coincide with the termination of neural inspiration. Ideally, the ventilator terminates inspiratory flow in synchrony with the patient's neural timing, but frequently the ventilator terminates inspiration either early or late. Most current mechanical ventilators include adjustable cycling features that, when used in conjunction with waveform graphics, can enhance patient-ventilator synchrony.