Brain Stimul
-
Randomized Controlled Trial Multicenter Study
Long-term measurement of impedance in chronically implanted depth and subdural electrodes during responsive neurostimulation in humans.
Long-term stability of the electrode-tissue interface may be required to maintain optimal neural recording with subdural and deep brain implants and to permit appropriate delivery of neuromodulation therapy. Although short-term changes in impedance at the electrode-tissue interface are known to occur, long-term changes in impedance have not previously been examined in detail in humans. To provide further information about short- and long-term impedance changes in chronically implanted electrodes, a dataset from 191 persons with medically intractable epilepsy participating in a trial of an investigational responsive neurostimulation device (the RNS(®) System, NeuroPace, Inc.) was reviewed. ⋯ Although there were statistically significant short-term impedance changes, long-term impedance was stable after one year. Impedances for depth electrodes transiently increased during the third week after lead implantation and impedances for subdural electrodes increased over 12 weeks post-implant, then were stable over the subsequent long-term follow-up. Both depth and subdural electrode impedances demonstrated long-term stability, suggesting that the quality of long-term electrographic recordings (the data used to control responsive brain stimulation) can be maintained over time.