Clin Chem Lab Med
-
During the West Africa Ebola virus disease (EVD) outbreak, a Belgian laboratory was deployed for supporting the Ebola treatment unit (ETU) of N'Zerekore, Guinea. Besides diagnosis of EVD and malaria, biochemical parameters were tested and used to guide supportive treatment of EVD. ⋯ POCT availability resulted in a paradigm shift in laboratory practice and care delivery at the N'Zerekore ETU. We conclude that there is urgent need for novel well-designed and validated POCT devices usable by non-expert operators in high ambient temperature and limited space. These devices should withstand regular and thorough decontamination by the personnel working on-site with life-threatening pathogens and be compatible with high biosafety level procedures. Such specific users' requirements need a European validation and standardization process of proposed solutions led by the EU Standardization Committee (CEN).