Is the international normalised ratio (INR) reliable? A trial of comparative measurements in hospital laboratory and primary care settings.
To determine the reliability of international normalised ratio (INR) measurement in primary care by practice nurses using near patient testing (NPT), in comparison with results obtained within hospital laboratories by varied methods. ⋯ No technical problems associated with INR testing within primary care were uncovered. Discrepant INR results are as problematic in hospital settings as they are in primary care. These data highlight the failings of the INR to standardise when different techniques and reagents are used, an issue which needs to be resolved. For primary care to become more involved in therapeutic oral anticoagulation monitoring, close links are needed between hospital laboratories and practices, particularly with regard to training and quality assurance.