Measuring tidal volume (VT) in nonintubated swine or swine with leaking breathing circuits is challenging. The aim of this study was to validate respiratory inductance plethysmography (RIP) for measuring VT in swine that are comparable in size to adult humans. To determine calibration curves, VT and RIP readings were obtained from anesthetized swine (n = 8; weight, 46-50 kg) during positive-pressure (mechanical) ventilation and spontaneous breathing. ⋯ The data revealed that VT was linearly correlated with the movements of both ribcage and abdomen as measured by using plethysmography over a large range of tidal volume (44 to 1065 mL). In addition, the intercept of the linear equation was small or even negative during spontaneous breathing but increased significantly (maximum, 145 mL, 59.2 ± 35.1 mL) during positive pressure ventilation. Our results indicate that VT in swine can be calculated by using a simple univariate linear regression equation with RIP readings obtained during either mechanical ventilation or spontaneous breathing.