Aust Crit Care
-
Comparative Study
Comparison of dynamic measurements of pulse contour with pulsed heat continuous cardiac output in postoperative cardiac surgical patients.
Cardiac output (CO) can be measured using bolus thermodilution via a pulmonary artery catheter (PAC) and as continuous cardiac output (CCO), using pulsed heat thermoditution. Pulse contour cardiac output (PCCO) measures continuous CO by analysis of the arterial waveform after calibration with thermodilution CO. The Pulsion Medical Systems (PiCCO system) achieves this by transpulmonary aortic thermodilution (TDtpa). ⋯ Whilst PCCO responded to rapid change, it developed significant error during haemodynmamic instability and requires frequent recalibration. CCO on the other hand has a considerable time lag in responding to changes in CO. The way a monitor measures CO must be taken into account when using the data in clinical management.
-
To assess the incidence of diarrhoea in intensive care patients with a length of stay (LOS) greater than 3 days who were receiving any type of enteral tube feeding, and to measure the effect of implementing a bowel management protocol. ⋯ Use of an evidence-based protocol, and improved monitoring and reporting of bowel activity, can decrease the incidence of diarrhoea in ICU patients.