Transfus Apher Sci
-
Comparative Study
Hemoglobin screening in prospective blood donors: comparison of different blood samples and different quantitative methods.
Portable photometers are increasingly used for hemoglobin screening in blood centers. In a total of 500 unselected prospective blood donors, venous hemoglobin concentrations were measured using a hematology analyzer, and fingerstick hemoglobin concentrations were measured on both a HemoCue B and a Biotest hemoglobin photometer. ⋯ The deviation exceeded the limit of +/-10g l(-1) in 9% of samples tested by HemoCue and in 20% of those investigated by Biotest. The use of earstick samples led to considerable overestimation of the actual hemoglobin concentration.