BMC anesthesiology
-
Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration.
Several techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the pulmonary artery catheter (PAC). However, these techniques usually require a central venous line, an additional catheter, or a special calibration procedure. A new arterial pressure-based cardiac output (COAP) device (FloTractrade mark, Vigileotrade mark; Edwards Lifesciences, Irvine, CA, USA) only requires access to the radial or femoral artery using a standard arterial catheter and does not need an external calibration. We validated this technique in critically ill patients in the intensive care unit (ICU) using COPAC as the method of reference. ⋯ Although the COAP algorithm shows a minimal bias with COPAC over a wide range of values in an inhomogeneous group of critically ill patients, the scattering of the data remains relative wide. Therefore, the used algorithm (V 1.03) failed to demonstrate an acceptable accuracy in comparison to the clinical standard of cardiac output determination.