Impedance cardiography for monitoring changes in cardiac output.
Impedance Cardiography (ICG) is a non invasive method useful for continuous monitoring of cardiac output but, it still has not found wide usage for measuring cardiac output in clinics and research. Most studies focused on comparing the cardiac output measured at rest with reference methods. In the present study we evaluated the validity of ICG against Doppler Echocardiography (DE) in measuring cardiac output changes that occur during static exercise. ⋯ The pooled means of cardiac output measured by ICG and DE do not differ significantly and bears a significant correlation (r = 0.812, P < 0.001). The bias (d +/- s) calculated is 0.15 +/- 0.64 L/min. ICG could provide valid information regarding the relative changes in cardiac output.