Randomized Controlled Trial Comparative Study Clinical TrialAccuracy of auscultatory blood pressure measurement with a long cuff.
Intra-arterial blood pressure was compared with simultaneous auscultatory measurements in 37 subjects with a wide range of blood pressures and arm circumferences; six cuffs of various lengths and widths were used. Nineteen subjects had an arm circumference of 34 cm or more (mean 40 cm) and the other 18 were considered to be non-obese and had a mean arm circumference of 30 cm. ⋯ The differences in error among measurements with the four largest cuffs in the obese group (13 X 30 cm, 14 X 30 cm, 14 X 38 cm, and a conical cuff) were clinically irrelevant, and there was even less to choose among all six cuffs in the non-obese subjects. These results suggest that auscultatory blood pressure may be measured with acceptable accuracy with a single long bladdered cuff both in subjects with large arms and in subjects with normal sized arms.