The assessment of oxygen saturation in arterial blood by pulse oximetry (SpO₂) is based on the different light absorption spectra for oxygenated and deoxygenated hemoglobin and the analysis of photoplethysmographic (PPG) signals acquired at two wavelengths. Commercial pulse oximeters use two wavelengths in the red and infrared regions which have different pathlengths and the relationship between the PPG-derived parameters and oxygen saturation in arterial blood is determined by means of an empirical calibration. This calibration results in an inherent error, and pulse oximetry thus has an error of about 4%, which is too high for some clinical problems. ⋯ SaO₂ was calculated by using each pair of PPG signals selected from the three wavelengths. In measurements on healthy subjects, SpO₂ values, obtained by the 780-808 nm wavelength pair were found to be in the normal range. The measurement of SpO₂ by two nearby wavelengths in the infrared with narrow line-width enables the assessment of SpO₂ without calibration.