Radiology
-
Purpose To assess the performance of hardware- and software-gating technologies in terms of qualitative and quantitative characteristics of respiratory motion in positron emission tomography (PET) imaging. Materials and Methods Between 2010 and 2013, 219 fluorine 18 fluorodeoxyglucose PET examinations were performed in 116 patients for assessment of pulmonary nodules. All patients provided informed consent in this institutional review board-approved study. ⋯ The mean increase ± standard deviation in lesion maximum standardized uptake value was 42.2% ± 38.9 between nongated and software-gated images, and lesion full width at half maximum values decreased by 9.9% ± 9.6. Conclusion Compared with vendor-supplied respiratory-gating hardware methods, software gating performed favorably, both qualitatively and quantitatively. Fully automated gating is a feasible approach to motion correction of PET images. (©) RSNA, 2016 Online supplemental material is available for this article.