Journal of diabetes science and technology
-
J Diabetes Sci Technol · Nov 2008
Blood glucose measurements in arterial blood of intensive care unit patients submitted to tight glycemic control: agreement between bedside tests.
Implementing tight glycemic control (TGC) in intensive care unit (ICU) patients requires accurate blood glucose (BG) monitoring. We evaluated the performance of two commercially available bedside glucometers, Accu-Chek and HemoCue, in patients admitted to the ICU and in whom TGC was applied. ⋯ When TGC is implemented in ICU practice, caution is warranted when adjusting insulin rates based only on BG readings obtained by the tested glucometers. ICU practitioners should weigh the advantages and disadvantages of such devices: a greater bias but with a more predictable error and measurement behavior versus a somewhat lower bias but with an unpredictable direction of the difference.