Journal of diabetes science and technology
-
J Diabetes Sci Technol · Mar 2015
The impact of measurement frequency on the domains of glycemic control in the critically ill--a Monte Carlo simulation.
The role of blood glucose (BG) measurement frequency on the domains of glycemic control is not well defined. This Monte Carlo mathematical simulation of glycemic control in a cohort of critically ill patients modeled sets of 100 patients with simulated BG-measuring devices having 5 levels of measurement imprecision, using 2 published insulin infusion protocols, for 200 hours, with 3 different BG-measurement intervals-15 minutes (Q15'), 1 hour (Q1h), and 2 hours (Q2h)-resulting in 1,100,000 BG measurements for 3000 simulated patients. The model varied insulin sensitivity, initial BG value and rate of gluconeogenesis. ⋯ Higher measurement frequency mitigated the deleterious effect of high measurement imprecision, defined as CV ≥ 15%. This Monte Carlo simulation suggests that glycemic control in critically ill patients is more optimal with a BG measurement interval no longer than 1h, with further benefit obtained with use of measurement interval of 15'. These findings have important implications for the development of glycemic control standards.