Comparative StudyBiomechanical evaluation of the phases during simulated endotracheal intubation (ETI): pilot study on the effect of different laryngoscopes.
Endotracheal Intubation (ETI) is a common airway procedure used to connect the larynx and the lungs through a windpipe in patients under emergency situations. The process is carried out by a laryngoscope inserted into the mouth, used to help doctors in visualizing the glottis and inserting the tube. Currently, very few studies on objective evaluation of the biomechanics of the doctors during the procedure have been done. ⋯ In this paper, the authors present a preliminary study on a methodology to objectively evaluate and segment the biomechanical performance of doctors during the ETI, using surface electromyography and inertial measurement units. In particular, the validation has been performed by comparing three kinds of laryngoscopes involving an expert doctor. Finally, results are presented and commented.
-
Conf Proc IEEE Eng Med Biol Soc · Jan 2013
Multivariate temporal symptomatic characterization of cardiac arrest.
We model the temporal symptomatic characteristics of 171 cardiac arrest patients in Intensive Care Units. The temporal and feature dependencies in the data are illustrated using a mixture of matrix normal distributions. We found that the cardiac arrest temporal signature is best summarized with six hours data prior to cardiac arrest events, and its statistical descriptions are significantly different from the measurements taken in the past two days. This matrix normal model can classify these patterns better than logistic regressions with lagged features.
-
Conf Proc IEEE Eng Med Biol Soc · Jan 2013
Development of a smart backboard system for real-time feedback during CPR chest compression on a soft back support surface.
The quality of cardiopulmonary resuscitation (CPR) is often inconsistent and frequently fails to meet recommended guidelines. One promising approach to address this problem is for clinicians to use an active feedback device during CPR. ⋯ Based on adult CPR manikin tests it was found that the accuracy of the estimated CC depth for a dual accelerometer feedback system is significantly better (7.3% vs. 24.4%) than for a single accelerometer system on soft back support surfaces, in the absence or presence of a backboard. In conclusion, the algorithm used was found to be suitable for a real-time, dual accelerometer CPR feedback application since it yielded reasonable accuracy in terms of CC depth estimation, even when used on a soft back support surface.
-
Conf Proc IEEE Eng Med Biol Soc · Jan 2013
A new method to increase the quality of cardiopulmonary resuscitation in hospital.
In order to ensure that high-quality cardio-pulmonary-resuscitation (CPR) is performed, many kinds of feedback devices have been developed that are helpful for achieving correct chest compression (CC) in manikin studies. However, the mattress compression depth (MCD) can cause overestimation of chest compression depth (CCD) during CPR using a feedback device. Herein, we propose a new method using a vinyl cover that encloses the foam mattress and is compressed by vacuum pump just before performing CPR, which could increase the performance of CCs during CPR.
-
Conf Proc IEEE Eng Med Biol Soc · Jan 2013
Demonstrating the accuracy of an in-hospital ambulatory patient monitoring solution in measuring respiratory rate.
This paper presents clinical testing conducted to evaluate the accuracy of Aingeal, a wireless in-hospital patient monitor, in measuring respiration rate via impedance pneumography. Healthy volunteers were invited to simultaneously wear a CE Marked Aingeal vital signs monitor and a capnograph, the current gold standard in respiration rate measurement. ⋯ Statistical analysis of the data collected shows a mean difference of -0.73, a standard deviation of 1.61, limits of agreement of -3.88 and +2.42 bpm and a P-value of 0.22. This testing demonstrates comparable performance of the Aingeal device in measuring respiration rate with a well-accepted and widely used alternative method.