Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi
-
Sheng Wu Yi Xue Gong Cheng Xue Za Zhi · Apr 2021
[Simulation study of spinal cord stimulation evoked compound action potential].
Spinal cord stimulation (SCS) for pain is usually implanted as an open loop system using unchanged parameters. To avoid the under and over stimulation caused by lead migration, evoked compound action potentials (ECAP) is used as feedback signal to change the stimulating parameters. This study established a simulation model of ECAP recording to investigate the relationship between ECAP component and dorsal column (DC) fiber recruitment. ⋯ The amplitude of this slow conduction crest increased as the stimulating intensity increased while the amplitude of the fast conduction crest almost remained unchanged. Therefore, the simulated ECAP signal in this paper could be used to evaluate the degree of excitation of DC fibers. This SCS-ECAP model may provide theoretical basis for future clinical application of close loop SCS base on ECAP.