American journal of hospital pharmacy
-
The contamination potential of sterile disposable plastic syringes was evaluated after subjecting the syringes to both simulated in-use conditions and an intentional microbial challenge. Lots of 20 Luer-lock syringes in 10 or 12-cu cm and 20- or 30-cu cm sizes from three manufacturers were tested. Sampling was conducted using 30-ml vials of sterile aerobic culture media containing 14C-labeled substrates. ⋯ When this sampling technique was applied to the syringes inoculated on the upper piston surface with Bacillus subtilis suspension, a 100% contamination rate was observed in 120 samples each under both positive and negative in-vial pressure. Grasping the ribbed plunger shaft of disposable plastic syringes with a dry bare hand did not compromise the sterility of the syringe contents in this study; however, this practice should be avoided when possible. Personnel should absolutely avoid introducing fluid-borne microbial contaminants into the distal barrel end of these syringes because the contents are readily labile to contamination under these conditions.