This study evaluated the efficacy of a new patient-triggered ventilator that triggered the patient's inspiratory effort by detecting the change in airflow by means of a 'hot wire' anemometer. This ventilator was used in both the conventional and the synchronized intermittent mandatory ventilation (SIMV) modes in seven neonates. Values for blood gas, spontaneous breathing rate, tidal volume of spontaneous breaths and minute volume were compared in all seven neonates. ⋯ The tidal volume of spontaneous breaths was more constant with SIMV versus conventional mechanical ventilation. Thus, the airway flow-triggered SIMV may lessen inspiratory muscle fatigue during weaning process. We conclude that the SIMV is useful in weaning neonates from the ventilator.