Zeitschrift für Kardiologie
-
Data on mid- and long-term follow-up for the recent devices for closure of secundum-type atrial septal defects are limited. The purpose of our retrospective study was to report the effectiveness of transcatheter closure in patients with various morphological types of atrial septal defect other than centrally located defects within the oval fossa using the CardioSEAL and CardioSEAL-Starflex occluder. A total of 91 patients (age 1.5-71 years, median 6 years) underwent transcatheter closure. ⋯ Serial transthoracic echocardiographic findings revealed protrusion of one left-sided arm onto the right atrial aspect in 5 patients; malposition of one right-sided superior arm of the device was observed in 7 patients. Fluoroscopy showed single fatigue fracture in 7 patients (7.7%) within the first 6 months after implantation. These results demonstrate that transcatheter closure with the double umbrella device was effective and safe on medium-term follow-up and could be extended to atrial septal defects of various morphology.