Ultrasound in medicine & biology
-
Ultrasound Med Biol · Jun 2003
3-D US frame positioning using speckle decorrelation and image registration.
In this paper, a new positioning system is proposed for the 3-D ultrasound (US). This system combines the image registration technique and speckle decorrelation algorithm to accurately position sequential ultrasonic images without any additional positioning hardware. The speckle decorrelation algorithm estimates the relative distance of two neighboring frames and the image registration technique gets the range of the whole 3-D ultrasonic data set and makes slight modification on each frame's position. ⋯ This reference image intersects each frame of the 3-D ultrasonic data set in a line. For each frame, the intersectional line is first found and then the location in the reference image can be used to estimate the position of this frame. This system uses the data set of consecutive 2-D freehand-scanned US B-mode images to construct the 3-D US volume data, and it can be integrated into the 3-D US volume rendering system.