Pulsatile flow phantom for ultrasound image-guided HIFU treatment of vascular injuries.
A pulsatile flow phantom was developed for studies of ultrasound image-guided high intensity focused ultrasound (HIFU) application in transcutaneous hemostasis of injured blood vessels. The flow phantom consisted of a pulsatile pump system with instrumented excised porcine carotid artery, which was imbedded in a transparent agarose gel to model structural configuration of in vivo tissues. Heparinized porcine blood was circulated through the phantom. ⋯ Hemostasis was achieved in 55 +/- 31 s (n = 9) of HIFU application. Histologic observations showed that a HIFU-sealed puncture was filled with clotted blood and covered with a fibrin cap. The pulsatile flow phantom provides a controlled and repeatable environment for studies of transcutaneous image-guided HIFU application in hemostasis of a variety of blood vessel injuries.