A rotating cylinder phantom for flow and tissue color Doppler testing.
Ultrasound Doppler using two-dimensional (2D) techniques is commonly used to study blood flow and myocardial tissue motion. This use includes measurement of velocity and time intervals, often in relation to the electrocardiogram (ECG) signal. 2D Doppler is frequently considered a real-time technique but in reality the acquisition time can be as long as 200 ms per image. ⋯ The phantom can produce constant velocities for velocity testing, as well as accelerating movement for testing the timing characteristics of ultrasound systems. Our investigation shows that the cylinder phantom is especially suitable for timing measurements in 2D Doppler imaging and that time delays between the Doppler signals and the ECG signal exist in the tested ultrasound system.