Ultrasound in medicine & biology
-
Ultrasound Med Biol · Feb 2020
Intra-observer and Device-Dependent Inter-observer Reliability of Contrast-Enhanced Ultrasound for Muscle Perfusion Quantification.
Muscle perfusion quantification by contrast-enhanced ultrasound (CEUS) may facilitate treatment decisions in musculoskeletal disorders. Translation into clinical routine relies on high intra-observer and inter-observer reliability and transferability between ultrasound devices to enable validation and multicenter studies. This study evaluates these aspects for deltoid muscle perfusion quantification, including possible multicenter study setups. ⋯ Re-evaluating subgroups with high sectional plane concordance significantly increased reliability (intra-observer: ICC = .97, CV = 5.49%, inter-observer/same device: ICC = .98, CV = 5.83%, varying devices: ICC = .78, CV = 9.8%). CEUS perfusion quantification of the deltoid seems applicable for multicenter studies, yet pooling different ultrasound devices remains critical. Sectional plane concordance appears to be crucial for reliability and transferability of CEUS muscle perfusion quantifications.