The purpose of the study described here was to investigate the reliability of impedance cardiography (IC) in measuring cardiac output (CO) and central blood volume. Absolute values and changes in these variables obtained by impedance cardiography and by isotope- or thermodilution techniques were compared. The reproducibility of IC within the same day was compared with that of isotope dilution and the reproducibility in IC from day to day was derived. ⋯ We conclude that impedance cardiography is reliable in measuring changes in cardiac output and thus suitable for repeated measurements in studies on the haemodynamic effects of physiological or pharmacological intervention. Impedance cardiography is sufficiently reliable for comparison of absolute values of CO between different groups of patients. We cannot recommend impedance cardiography for quantitative studies of central blood volume.