Hand (New York, N.Y.)
-
Hand (New York, N.Y.) · Sep 2007
Case ReportsIntraoperative fracture of a pyrocarbon PIP total joint-a case report.
Pain or dysfunction of the finger joints due to arthritis or traumatic injuries that fail medical management may necessitate arthroplasty or joint replacement. The goals of the finger joint implant arthroplasty are to relieve pain, to correct deformity, and to improve the function and appearance of the hand. Several prosthetic implants have been used for the replacement of the proximal interphalangeal (PIP) joint. ⋯ The component was removed, and a new implant was inserted without complication. Critical evaluation revealed that there was an inadequate resection of the volar lip resulting in a stress riser on the implant during impaction. Careful attention to this technical point will hopefully minimize the occurrence of this complication as the availability and use of these implants increases.