Thorax
-
Data on the effect on mucociliary clearance of oral high frequency oscillation is conflicting. By means of a technique to superimpose high frequency oscillation on tidal breathing, changes in mucociliary clearance during high frequency oscillation were studied in seven normal non-smokers by monitoring the clearance of inhaled radiolabelled aerosol from the lungs. After inhalation of 5 microns technetium 99m labelled particles under controlled conditions, whole lung clearance was monitored by scintillation counters half hourly for six hours with a final count at 24 hours, from which tracheobronchial deposition and clearance could be calculated. ⋯ Between 3 and 4.5 hours mucociliary clearance with high frequency oscillation exceeded control by about 10% (p less than 0.05). The mean time taken to eliminate 90% of deposited radioaerosol from the tracheobronchial tree fell from 4 hours 50 minutes (range 1 h 52 min-6 h 50 min) during control to 3 hours 43 minutes (range 2 hr 28 min-5 hr 54 min) during the high frequency oscillation run (p less than 0.05). Possibly this comfortable, simple technique would be of therapeutic benefit to patients with chronic sputum retention and merits further investigation.