Haematologica
-
There are Council of Europe recommendations for the quality of blood components. We analyzed the quality of blood components processed by a top & bottom system (Optipress((R)) II), the routine method used in our blood bank, to test whether the components reached the recommended quality. ⋯ Optipress II((R)) provides standardized, leukocyte-poor blood components. Council of Europe requirements were met in a large percentage of red-cell concentrates, with less than 92 and 74% of the original platelets and leukocytes, respectively, and a small hemoglobin loss per unit. The system gave an optimal yield in terms of plasma volume. The top & bottom technique allowed us to reduce the number of blood units per platelet concentrate from 6 to 4 units, with similar platelet yields compared with traditional procedures. Nevertheless, the storage conditions must be improved to satisfy all Council of Europe requirements for platelet concentrates.