Annales de l'anesthésiologie française
-
A new automatic apparatus for the measurement of arterial pressure by a non-invasive technique was compared with direct intra-arterial measurement in 20 adult patients in a surgical intensive care unit. The apparatus works on the basis of the principle of oscillometry. Blood pressure is determined with a microprocessor by analysis of the amplitude of the oscillations produced by a cuff which is inflated then deflated automatically. ⋯ These results indicate a good degree of agreement for measurements of mean arterial pressure, clinically the most important, between the two methods used. Measurements of diastolic pressure and above all of diastolic pressure seemed to be less in agreement. This difference could be due to an error in determination of the automatic apparatus tested or to the peripheral site (radial artery) of the intra-arterial catheter used, itself falsifying the humeral arterial pressure.
-