Clinical orthopaedics and related research
-
Clin. Orthop. Relat. Res. · Jul 2001
Randomized Controlled Trial Multicenter Study Comparative Study Clinical TrialFixation strength and pin tract infection of hydroxyapatite-coated tapered pins.
In a multicenter, prospective, randomized study, the biomechanical and clinical properties of the bone-pin interface were compared with standard tapered pins and hydroxyapatite-coated tapered pins implanted in patients who underwent femoral and tibial external fixation treatments. The results showed that the hydroxyapatite-coated tapered pins are clinically effective in improving the strength of fixation of the bonepin interface. This improvement corresponded to a lower rate of pin tract infection. ⋯ In this pin group, the mean pin extraction torque was 531 +/- 225 Ncm in the infected pin tracts and 508 +/- 233 Ncm in the uninfected pin tracts. In the standard pin group, the mean pin extraction torque was 73 +/- 142 Ncm in the infected pin tracts and 211 +/- 216 Ncm in the uninfected pin tracts. The advantages provided by the hydroxyapatite-coated pins were higher in cancellous bone than in cortical bone.