Bulletin européen de physiopathologie respiratoire
-
Bull Eur Physiopathol Respir · Jul 1986
Air entrainment during high-frequency jet-ventilation. Simulation of a bronchoscopy with a lung model.
Air entrainment contribution to jet-ventilation during bronchoscopy was evaluated as a lung model (increasing compliance and airway resistance). Ventilation was provided through a 10 mm internal diameter tube using either jet alone without air entrainment, or injection with air entrainment (coaxial and lateral injectors). Three I/E ratios (0.25, 0.43 and 0.67) and nine rates of ventilation, ranging from 20 to 300 c X min-1, were assessed. ⋯ CO2 elimination is related to the magnitude of ventilated volumes. The amount of entrained air interferes with the FIO2 of delivered gases. During bronchoscopy, lateral injection should be preferred because of smaller airway pressures (- 31.2 +/- 0.6%) and lung volumes, while tidal volumes remain adequate.