Bone & joint research
-
Bone & joint research · Jun 2018
Biomechanical evaluation of a new gliding screw concept for the fixation of proximal humeral fractures.
Plating displaced proximal humeral fractures is associated with a high rate of screw perforation. Dynamization of the proximal screws might prevent these complications. The aim of this study was to develop and evaluate a new gliding screw concept for plating proximal humeral fractures biomechanically. ⋯ Biomechanically, plate fixation using a new gliding screw technology did not show considerable advantages in comparison with fixation using a standard PHILOS plate. Based on the finding of telescoping of screws, however, it may represent a valid approach for further investigations into how to avoid the cut-out of screws.Cite this article: Y. P. Acklin, I. Zderic, J. A. Inzana, S. Grechenig, R. Schwyn, R. G. Richards, B. Gueorguiev. Biomechanical evaluation of a new gliding screw concept for the fixation of proximal humeral fractures. Bone Joint Res 2018;7:422-429. DOI: 10.1302/2046-3758.76.BJR-2017-0356.R1.