Electromyography and clinical neurophysiology
-
Electromyogr Clin Neurophysiol · Apr 1997
Evaluation of diaphragm electromyogram contamination during progressive inspiratory maneuvers in humans.
The diaphragm electromyogram (EMGdi) is susceptible to contamination by non-diaphragm related electrical signals such as the ECG, electrode motion artifacts, and other sources of noise. It is difficult to distinguish between these contaminating signals and those that are representative of the non-contaminated EMGdi, especially during periods when the EMGdi amplitude is relatively small, as during mild contractions of the diaphragm. The aim of the present study was to evaluate how contaminating signals influence the EMGdi power spectrum center frequency (CF) during progressive inspiratory maneuvers. ⋯ This was expressed by a factor of 4 reduction in the coefficient of variation of the CF values. The majority of the excluded EMGdi signals (i.e. not satisfying the spectral deformation index inclusion levels), had low CF values mainly due to the presence of electrode motion artifacts. It was concluded that: 1) The majority of EMGdi power spectrums are deformed early on during unloaded inspirations, and their CF values should be carefully interpreted as being representative of diaphragm function. 2) The relative contribution of contaminating signals in the EMGdi decreases proportionally throughout the first two thirds of an inspiration to IC. 3) The use of visual inspection of the signal in the time domain is questionable as a method to discriminate non-contaminated signals. 4) Analysis of the signal in the frequency domain makes it possible to detect the influence of signal contamination.