Pediatric research
-
Lung deposition of inhaled drugs in ventilated neonates has been studied in models of questionable relevance. With conventional nebulizers, pulmonary deposition has been limited to 1% of the total dose. The objective of this study was to assess lung delivery of aerosols in a model of neonatal ventilation using a conventional and novel electronic micropump nebulizer. ⋯ Duration of delivery was shorter with APN-C than with the two other nebulizers (2 versus 6 and 10 min for the APN-S and the MistyNeb, respectively; p < 0.001). Electronic micropump nebulizers are more efficient to administer aerosols in an animal model of ventilated neonates. Availability of Aerogen's electronic micropump nebulizers offers new opportunities to study clinical efficacy and risks of aerosol therapy in ventilated neonates.