Neurally adjusted ventilatory assist (NAVA), a mode of mechanical ventilation controlled by diaphragmatic electrical activity (EAdi), may improve patient-ventilator interaction. We examined patient-ventilator interaction by comparing EAdi to ventilator pressure during conventional ventilation (CV) and NAVA delivered invasively and non-invasively. Seven intubated infants [birth weight 936 g (range, 676-1266 g); gestational age 26 wk (range, 25-29)] were studied before and after extubation, initially during CV and then NAVA. ⋯ Pressure delivery during conventional ventilation was not correlated to EAdi. Neural expiratory time was longer (p = 0.044), and respiratory rate was lower (p = 0.004) during NAVA. We conclude that in low birth weight infants, NAVA can improve patient-ventilator interaction, even in the presence of large leaks.