Canadian journal of respiratory therapy : CJRT = Revue canadienne de la thérapie respiratoire : RCTR
-
Oscillatory positive expiratory pressure (OPEP) devices facilitate secretion clearance by generating positive end expiratory pressure. However, different device designs may produce different levels of expiratory pressure with the same expiratory flow rate. We bench tested four devices to determine the relationship between expiratory flow and expiratory pressure in each. ⋯ In this bench test of OPEP devices, we found considerable variability in expiratory flow requirements needed to generate an expiratory pressure of >10 cm H2O. Our finding suggests that smaller patients or those with limited expiratory airflow due to diseases such as COPD, obesity, chronic congestive heart failure, and restrictive lung disease may have better results when matched to OPEP devices requiring less expiratory airflow.