Archivos de cardiología de México
-
Hemodynamic monitoring has been used extensively during the last decades for risk stratification and guiding treatment of patients with cardiovascular destabilization, especially in the scenario of acute heart failure and cardiac shock. Every cardiac pump has its own maximum performance, which denotes its pumping capability. The heart is a muscular mechanical pump with an ability to generate both flow (cardiac output) and pressure. ⋯ Others investigators observed cutoff for increased mortality of CP < 1 W, data that were obtained at doses of maximal pharmacologic support yielding the individual maximal CP. In our experience, the cutoff value for CP that accurately predicts in-hospital mortality for cardiogenic shock patients is 0.7 W, but its impact on short-term prognosis is clearer if the patient achieves a CP equal or higher than 1 W after an optimal myocardial revascularization with interventional cardiac procedures. According to the data collected from the literature, CP deserves a place in the evaluation of the patient with cardiogenic shock due to an acute myocardial infarction, but a more profound analysis of this parameter an further evaluation are required in order to better understand its prognostic meaning in this acute cardiac syndrome.