Randomized Controlled Trial Multicenter Study Comparative StudyGlycated hemoglobin assessment in clinical practice: comparison of the A1cNow point-of-care device with central laboratory testing (GOAL A1C Study).
The Glycemic Optimization with Algorithms and Labs At Po1nt of Care (GOAL A1C) Study assessed the effect of titration monitoring strategies and methods of A1C testing on glycemic control in patients with type 2 diabetes failing oral therapy and beginning basal insulin glargine. The availability of both point-of-care (POC) and central laboratory A1C values provided an opportunity to evaluate correlation and statistical agreement between these methods of testing. This analysis forms the basis of the current report. ⋯ POC testing of A1C in predominantly primary care settings using the A1cNow device was correlated with central laboratory results. The correlation was less than expected based on each method's reproducibility data. Although there was agreement between the average POC A1C values and the corresponding central laboratory values, the dispersion of individual POC A1C values was large. Thus, we conclude that these two methods of A1C testing should not be used interchangeably.