Journal of acquired immune deficiency syndromes : JAIDS
-
J. Acquir. Immune Defic. Syndr. · Sep 2010
Comparative StudyEvaluation of the PIMA point-of-care CD4 analyzer in VCT clinics in Zimbabwe.
Point-of-care (POC) CD4 testing was implemented at a stand-alone HIV voluntary testing and counseling centre in Harare, Zimbabwe. To validate the use of this new technology, paired blood samples were collected from 165 patients either by a nurse or a laboratory technician and tested using POC and conventional laboratory CD4 machines. Finger prick (capillary) blood was collected directly into the PIMA POC CD4 Analyzer cartridges and tested immediately, whereas venous blood collected into evacuated tubes was used for CD4 enumeration on a Becton Dickinson FACSCalibur. ⋯ Both nurses and laboratory technicians performed the test accurately, thereby increasing the human resources available for POC CD4 testing. By producing same-day results, POC CD4 facilitates immediate decision-making, patient management and referral and may help improve patient care and retention. POC CD4 may also alleviate testing burdens at traditional central CD4 laboratories, hence improving test access in both rural and urban environments.