Efficiency in endotoxin removal by a reprocessing protocol for electrophysiology catheters based on hydrogen peroxide plasma sterilization.
Electrophysiology and ablation cardiac catheters, which come in contact with blood during clinical use, are required to be non-pyrogenic (<20 endotoxin units (EU)/device). This study aimed to quantify the residual endotoxin load in reprocessed devices as a mandatory step to guarantee safe reuse. We monitored the pyrogenic status of the device (n=61) in three fundamental steps of the reprocessing protocol: after clinical use, after decontamination-cleaning treatments and after complete reprocessing, including sterilization by hydrogen peroxide gas plasma. ⋯ Depyrogenation testing demonstrated efficiency in endotoxin reduction of more than 62 times (1.8log). These results show the determining role of hydrogen peroxide gas-plasma sterilization in the reduction of pyrogenic load on medical devices. Considering actual hygienic requirements at single-use device reprocessing, hydrogen peroxide gas-plasma sterilization can be considered as an efficient treatment at non-lumen cardiac electrophysiology catheter reprocessing.