The handheld blood lactate analyser versus the blood gas based analyser for measurement of serum lactate and its prognostic significance in severe sepsis.
This study was done to compare the accuracy of the Lactate Pro LT 1710 (Arkray Inc., Kyoto, Japan) with the Combiline Plus (Eschweiler GmbH & Co. KG Holzkoppelweg, Kiel, Germany), and also, to analyze the prognostic significance of serum lactates and Simplified Acute Physiology Score 3 (SAPS 3) in patients of severe sepsis. ⋯ The Lactate Pro LT 1710 provides similar results to the Combiline Eschweiler blood gas analyser and is a cheaper alternative. It would prove to be a boon in peripheral hospitals in the aggressive management of critically ill patients.