Pediatric diabetes
-
Randomized Controlled Trial Multicenter Study Comparative Study Clinical Trial
The "Glucositter" overnight automated closed loop system for type 1 diabetes: a randomized crossover trial.
Tight glucose control is needed to prevent long-term diabetes complications; this is hindered by the risk of hypoglycemia, especially at night. ⋯ The results of this study demonstrate the ability of the MDLAP to safely improve overnight glucose control without increased risk of hypoglycemia in patients with T1DM at three different national, geographic, and clinical centers (ClinicalTrials.gov number, NCT 01238406).