Applied occupational and environmental hygiene
-
Appl Occup Environ Hyg · Oct 2002
Comparative StudyComparison of five methods for fit-testing N95 filtering-facepiece respirators.
Five fit-testing methods (Bitrex, ambient aerosol condensation nuclei counter using the TSI PortaCount Plus, saccharin, modified ambient aerosol condensation nuclei counter using the TSI PortaCount Plus with the N95-Companion, and generated aerosol using corn oil) were evaluated for their ability to identify poorly fitting N95 filtering-facepiece respirators. Eighteen models of NIOSH-certified, N95 filtering-facepiece respirators were tested by a panel of 25 subjects using each fit-testing method. The penetration of the corn oil and the ambient aerosols through the filter media of each respirator was measured in order to adjust the corresponding generated and ambient aerosol overall fit factors, reflecting only face-seal leakage. ⋯ The ambient aerosol using the TSI PortaCount Plus and the generated aerosol methods identified poorly fitting respirators better than the saccharin, the Companion, and Bitrex methods. These errors rates should be considered when selecting a fit-testing method for fitting N95 filtering-facepieces. When both types of errors were combined as an assignment error, the ambient aerosol method using the TSI PortaCount Plus had the lowest percentage of wearers being assigned a poor-fitting respirator.