Hyperpolarized noble gas (HNG) magnetic resonance imaging (MRI) has been shown to be useful for studying rodent models of lung disease. Image quality can be substantially degraded by signal loss from molecular oxygen entering the airway, requiring invasive surgery to ensure a good seal between the endotracheal (ET) tube and trachea. A modified Foley catheter having an inflatable cuff near the tip provides a novel approach for ensuring image quality for HNG MRI, thereby enabling longitudinal studies and reducing animal numbers. ⋯ The Foley catheter resulted in minimal leak rates (-0.20 (0.03) versus -0.16 (0.05) cmH(2)O/s), and were shown to be well below upper-limit leak rates of -0.5 and -0.7 cmH(2)O/s. Tracheal samples from rats in a separate Foley group (not mechanically ventilated) showed superficial damage six days postextubation (grade = 0). (3)He imaging performed using the Foley showed good image quality. Though some technical issues remain to be solved, a modified Foley catheter used as an ET tube offers the potential to enable longitudinal studies in rodents and reduce animal numbers.